Skip to main navigation Skip to search Skip to main content

Assembly of a polymer lab-on-chip device for impedimetric measurements of D-dimers in whole blood

  • A. Ohlander
  • , C. Strohhoefer
  • , K. Bock
  • , S. M. Scott
  • , Zulfiqur Ali
  • , Peter Musil
  • , Jan Kyselovic
  • Fraunhofer Gesellschaft
  • Technical University of Berlin
  • University of Teesside
  • Comenius University Bratislava

Research output: Chapter in Book/Report/Conference proceeding › Conference contribution › peer-review

1 Citation (Web of Science)

Abstract

This paper reports the development and characterisation of an assembly technology for a polymer lab-on-chip. The system consists of a 150 mu m deep hot embossed microfluidic channel in polycarbonate and Au electrodes fabricated separately by photolithography on polyethylenenaphthalate. The system is designed for impedimetric immunoassay detection in whole blood. Electrode layer and microfluidic substrate are joined by means of a 50 mu m thick double-sided medical grade adhesive tape, adjusted with an optical alignment system. The bond proved to be liquid tight at room temperature. An alignment accuracy of 34 mu m (+/-19 mu m) evaluated over a set of 23 samples, was achieved. The effect of alignment accuracy of the intermediate adhesive film on whole blood flow properties in the device is studied. Already an alignment error of 70 mu m increases the flushing out time of whole blood by approximately 20 %.
Original languageEnglish
Title of host publication2010 Proceedings 60th Electronic Components and Technology Conference (ECTC)
PublisherIEEE
Pages1004-1009
Number of pages6
ISBN (Electronic)978-1-4244-6412-8, 978-1-4244-6411-1
ISBN (Print)978-1-4244-6410-4
DOIs
Publication statusPublished - 21 Jun 2010
EventElectronic Components and Technology Conference - Las Vegas, United States
Duration: 1 Jun 2010 → 4 Jun 2010
Conference number: 60th
https://ectc.net/

Publication series

NameProceedings Electronic Components and Technology Conference (ECTC)
PublisherIEEE
Number60th
Volume2010
ISSN (Print)0569-5503
ISSN (Electronic)2377-5726

Conference

ConferenceElectronic Components and Technology Conference
Abbreviated titleECTC 2010
Country/TerritoryUnited States
CityLas Vegas
Period1/06/10 → 4/06/10
Internet address

Fingerprint

Dive into the research topics of 'Assembly of a polymer lab-on-chip device for impedimetric measurements of D-dimers in whole blood'. Together they form a unique fingerprint.

Cite this